• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: INTUITIVE SURGICAL, INC None; HARMONIC ACE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

INTUITIVE SURGICAL, INC None; HARMONIC ACE Back to Search Results
Model Number 480275-10
Medical Device Problem Code Thermal Decomposition of Device (1071)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/07/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during a da vinci-assisted cholecystectomy surgical procedure, the tip of the harmonic ace seemed to melt during use.Customer had to get a new handpiece and cord.Customer continued the procedure as expected.The procedure was completed with no reported injury.
 
Additional Manufacturer Narrative
Intuitive surgical, inc.(isi) received the da vinci product to perform failure analysis.The harmonic ace instrument was analyzed and found to have teflon pad damage.The teflon pad was damaged at the tip of the instrument, and the damage is axially aligned with the pad.The teflon pad damaged, and there may be missing pieces; however, the size of the missing pieces cannot be confirmed, indicating that a fragment may have detached from the instrument.The probable root cause of the teflon pad damage is attributed to use conditions.Damage is caused by heat generation when activating the harmonic ace blade with little to no tissue between the pad and the blade itself.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
None
Common Device Name
HARMONIC ACE
Manufacturer (Section D)
INTUITIVE SURGICAL, INC
3410 central expressway
santa clara CA 95051
Manufacturer (Section G)
INTUITIVE SURGICAL, INC
3410 central expressway
santa clara CA 95051
Manufacturer Contact
izabel nielson
3410 central expressway
santa clara, CA 95051
+1(408)523-2100
MDR Report Key50310175
Report Number2955842-2026-36058
Device Sequence Number8770427
Product Code NAY
UDI-Device Identifier10886874112738
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K143132
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other
Remedial Action Other
Type of Report Initial
Report Date (Section B) 08/18/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number480275-10
Device Catalogue Number480275
Device Lot NumberK11260226 0283
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer07/25/2026
Is the Reporter a Health Professional? No
Initial Date Received by Manufacturer 07/27/2026
Initial Report FDA Received Date08/18/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/26/2026
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
da Vinci Instruments and Accessories
Patient Age73 YR
Patient SexFemale
Patient Weight138 LB
Patient RaceWhite
-
-