| Model Number |
480275-10 |
| Medical Device Problem Code |
Thermal Decomposition of Device (1071)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/07/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during a da vinci-assisted cholecystectomy surgical procedure, the tip of the harmonic ace seemed to melt during use.Customer had to get a new handpiece and cord.Customer continued the procedure as expected.The procedure was completed with no reported injury.
|
| |
|
Additional Manufacturer Narrative
|
|
Intuitive surgical, inc.(isi) received the da vinci product to perform failure analysis.The harmonic ace instrument was analyzed and found to have teflon pad damage.The teflon pad was damaged at the tip of the instrument, and the damage is axially aligned with the pad.The teflon pad damaged, and there may be missing pieces; however, the size of the missing pieces cannot be confirmed, indicating that a fragment may have detached from the instrument.The probable root cause of the teflon pad damage is attributed to use conditions.Damage is caused by heat generation when activating the harmonic ace blade with little to no tissue between the pad and the blade itself.
|
| |
|