| Model Number |
470057-11 |
| Medical Device Problem Code |
Image Orientation Incorrect (1305)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/24/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during a da vinci assisted incisional hernia etep surgical procedure, the customer informed the technical support engineer (tse) that they could not tell if the endoscope orientation appeared correct.The customer replaced the endoscope, but the issue did not change.The tse then reviewed the logs and found no related errors, and the tse walked the customer through removing the endoscope from arm 3, orienting it to the desired position, pressing the arm 3 instrument clutch button twice to cancel guided tool change, and reinstalling the scope, after which the customer reported the problem did not persist.The procedure was completing as planned.No known impact or patient consequence was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
An investigation was completed to determine the cause of this reported event.The reported event was addressed with phone support.The technical support engineer (tse) guided the user to remove the endoscope from arm 3, reposition it, press the arm 3 instrument clutch button twice to cancel guided tool change, and reinstall the scope; after, the customer reported the issue was resolved and did not recur.No site visit was conducted.The system was working properly, and no additional action was required as the issue was resolved.The probable root cause is attributed to improper setup.
|
| |
|