| Model Number |
L300 |
| Medical Device Problem Code |
Premature Discharge of Battery (1057)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/09/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that this pacemaker recorded a code 1003, indicating the voltage is too low for the projected remaining capacity.A request was made to have data from this device analyzed.Data analysis identified potential high impedance within the battery.The pacemaker will remain implanted and the patient will be monitored until radio frequency (rf) telemetry disablement, at which time device replacement should be scheduled.No adverse patient effects were reported.
|
| |
|
Event or Problem Description
|
|
It was reported that this pacemaker recorded a code 1003, indicating the voltage is too low for the projected remaining capacity.A request was made to have data from this device analyzed.Data analysis identified potential high impedance within the battery.The pacemaker will remain implanted and the patient will be monitored until radio frequency (rf) telemetry disablement, at which time device replacement should be scheduled.No adverse patient effects were reported.Additional information received reported that this pacemaker was explanted and replaced.No additional adverse patient effects were reported.Product return was requested.
|
| |
|
Additional Manufacturer Narrative
|
|
The local area field representative was contacted for additional information regarding product return status.At this time, no further information is available.Should additional information become available this report will be updated.This product investigation is still in progress.This report will be updated when product investigation is complete.
|
| |
|
Additional Manufacturer Narrative
|
|
The local area field representative was contacted for additional information regarding product return status.At this time, no further information is available.Should additional information become available this report will be updated.Correction to h1: report type updated to serious injury.Correction to h6: impact code f26 removed; impact codes f1905 and f08 added to reflect device changeout.This product investigation is still in progress.This report will be updated when product investigation is complete.
|
| |
|