| Model Number |
E162 |
| Medical Device Problem Codes |
Premature Discharge of Battery (1057); Delayed Charge Time (2586); Battery Problem (2885)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/22/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that this implantable cardioverter defibrillator (icd) was suspected of premature battery depletion (pbd) as the battery has depleted from 7 months down to reaching elective replacement indicator state.The boston scientific technical services consultant evaluated the transmission data which demonstrated a charge time of 15.7 seconds, representing a secondary criterion for recommending a device replacement while the system continues to operate normally.Subsequently, this icd was explanted and replaced.No additional adverse patient effects were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
If pertinent information is provided in the future, a supplemental report will be submitted.
|
| |
|
Search Alerts/Recalls
|