BOSTON SCIENTIFIC CORPORATION INCEPTA? ICD; implantable cardioverter defibrillator Back to Search Results
Model Number E162
Medical Device Problem Codes Premature Discharge of Battery (1057); Delayed Charge Time (2586); Battery Problem (2885)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/22/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that this implantable cardioverter defibrillator (icd) was suspected of premature battery depletion (pbd) as the battery has depleted from 7 months down to reaching elective replacement indicator state.The boston scientific technical services consultant evaluated the transmission data which demonstrated a charge time of 15.7 seconds, representing a secondary criterion for recommending a device replacement while the system continues to operate normally.Subsequently, this icd was explanted and replaced.No additional adverse patient effects were reported.
 
Additional Manufacturer Narrative
If pertinent information is provided in the future, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INCEPTA? ICD
Common Device Name
implantable cardioverter defibrillator
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
Boston Scientific Corporation
cashel road
clonmel E91 T 862
EI   E91 T862
Manufacturer Contact
mark mischke
4100 hamline avenue north
saint paul, MN 55112
+1(612)875-7234
MDR Report Key50313817
Report Number2124215-2026-44520
Device Sequence Number9982819
Product Code LWP
UDI-Device Identifier00802526480775
UDI-Public(01)00802526480775(17)160821
Combination Product (Y/N)N
Initial Reporter StateVA
Initial Reporter CountryUS
PMA/510(K) Number
P960040
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2014
Device Explanted Year2026
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 09/29/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date08/21/2016
Device Model NumberE162
Device Catalogue NumberE162
Device Lot Number113311
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer Not provided
Supplement Date Received by ManufacturerNot provided
09/28/2026
Initial Report FDA Received Date08/19/2026
Supplement Report FDA Received Date09/03/2026
09/29/2026
Date Device Manufactured08/28/2014
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient Age84 YR
Patient SexMale
-
-