Model Number 1888TC/46
Medical Device Problem Codes Break (1069); Failure to Capture (1081); High impedance (1291)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/30/2026
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the atrial lead exhibited high pacing impedance and failed to capture when pacing at high outputs.During the procedure, the physician observed evidence of insulation damage on the atrial lead.When attempting to reconnect the existing atrial lead to the replacement pulse generator, high impedance was again noted on the atrial lead, and capture could not be achieved even at high pacing outputs.Subsequently, the atrial lead was capped on (b)(6) 2026 and replaced.The patient was stable throughout.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TENDRIL ST
Common Device Name
Drug Eluting Permanent Right Ventricular (Rv) Or Right Atrial (Ra) Pacemaker Electrodes
Manufacturer (Section D)
Abbott Medical (CRM-Sylmar)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
Abbott Medical (CRM-Sylmar)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
navpreet singh
15900 valley view court
sylmar, CA 91342
MDR Report Key50315829
Report Number2017865-2026-17411
Device Sequence Number9674941
Product Code NVN
UDI-Device Identifier05414734501743
UDI-Public(01)05414734501743(10)2763386(17)120229
Combination Product (Y/N)Y
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
P960013
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2009
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 08/19/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date02/29/2012
Device Model Number1888TC/46
Device Lot Number2763386
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date08/19/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured02/18/2009
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
7121/60 DURATA.; Ellipse DR ICD.
Outcome Attributed to Adverse Event Required Intervention;
Patient Age79 YR
Patient SexFemale
-
-