CAREFUSION 303, INC. BD Pyxis¿ MedBank Mini; Automated Dispensing Cabinet Back to Search Results
Model Number 500004500500
Medical Device Problem Code Thermal Decomposition of Device (1071)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/04/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that when using the bd pyxis¿ medbank mini, the drawer failed to open.As part of the investigation, it was determined that an excessive current on the circuit board.The excessive current thermally damaged u501 (voltage regulator).There were no adverse events or injuries reported based on this event.
 
Additional Manufacturer Narrative
A review of the complaint history for sn (b)(6) was performed in salesforce which did not locate similar complaint(s) with the same failure mode for this serial number.A review of the device history record for sn (b)(6) was performed from the date of manufacture, 26-apr-2021 and confirmed that this device was not previously returned for servicing and there were no production failures which correlates to the customer reported issue.The reported condition of drawer failure was confirmed during field service engineer (fse) testing and subsequently confirmed in the dchu inspection process.According to work order (b)(4), to resolve the drawer issue, field service engineer (fse) replaced pyxibus module controller (pmc) board on full height drawer 1.During dchu visual inspection: part number (p/n) 151730-21 or part name pcba pmc pyxis mini fw1-12: the returned pmc mini drawer board was externally inspected for any irregularities.U501 (regulator ic) was observed thermally damaged.There were no other anomalies observed.During dchu testing: part number (p/n) 151730-21 or part name pcba pmc pyxis mini fw1-12: due to the thermal damage observed on u501 further testing could not be performed.The device was in use for treatment purposes as intended per 21 cfr 820.198(d).Root cause: the root cause of the reported drawer failure was identified and attributed to an excessive current on the circuit board.The excessive current thermally damaged u501 (voltage regulator).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD Pyxis¿ MedBank Mini
Common Device Name
Automated Dispensing Cabinet
Manufacturer (Section D)
CAREFUSION 303, INC.
3750 torrey view ct
san diego CA 92130
Manufacturer (Section G)
CareFusion 303, INC.
10020 pacific mesa blvd
san diego CA 92121 -438
Manufacturer Contact
brett wilko
3750 torrey view ct
san diego, CA 92130
+1(858)617-2000
MDR Report Key50315930
Report Number2016493-2026-56132
Device Sequence Number18654020
Product Code BRY
UDI-Device Identifier10885403512582
UDI-Public(01)10885403512582
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 08/17/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model Number500004500500
Device Catalogue Number000000000000000349
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer02/06/2026
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date08/19/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/26/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1