• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Unomedical Devices S.A de C.V Inset; Set, Administration, Intravascular

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Unomedical Devices S.A de C.V Inset; Set, Administration, Intravascular Back to Search Results
Model Number 1001682
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code Hyperglycemia (1905)
Date of Event 02/28/2025
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the patient faced infusion set tubing was leaking at the site event on (b)(6) 2025 and also experienced high blood glucose levels which was treated by correction injection via mdi.
 
Additional Manufacturer Narrative
Initial 1 of 2.Unomedical hereby submits this initial report as part of its complaint remediation activities conducted under a capa 2356421 and capa 2566479 in accordance with the fda action plan, which incorporates qp-ic-2026-0020 (ic retrospective complaint review) and qp-ic-2026-0024 (ic retrospective mdr review).This submission includes a retrospective reassessment of previously evaluated complaints.Unomedical is providing this supplemental information in accordance with the reporting requirements set forth in 21 cfr part 803.Any fields left blank indicate that the information is unknown, unavailable, or remains unchanged additional information - this medical device report (mdr)is being submitted to include the below: h6: investigation results under type of investigation, investigation findings, investigation conclusions.H11: investigation summary.Conclusion summary of complaint investigation: a comprehensive review was conducted, including eqms queries, similar complaint searches, medical device file review, visual evidence assessment, and capa determination.No ncrs or capas of the same or similar nature were found for lot 6008099 and related malfunction codes.Two complaints was identified for this lot however, no trend or systemic issue was detected.The manufacturing records confirmed that the lot was produced in compliance with all requirements, with no deviations or maintenance events noted.Photos were requested; however, the customer confirmed that no photos were available for analysis.Consequently, an investigation was conducted using reference samples, and no failures related to the complaint were identified.No photo evidence was provided, so the assessment was based on documentation and reference sample analysis only.Based on these results, no manufacturing or quality issues were identified.The record will be closed and monitored through routine tracking and trending.For more details, see the information under the child investigation record (b)(4).After assessment of the failure via product trending over time with control charts, the risk has been deemed as within accepted limits.No further action is required.Complaint unconfirmed: following the investigation, the complaint is unconfirmed as analyzed.Based on the available information, this event is deemed to be a serious injury to date no additional information has been received.Should additional information become available, a follow-up report will be submitted.Fda registration number reporting site: 8021545.Manufacturing site: 3003442380.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Inset
Common Device Name
Set, Administration, Intravascular
Manufacturer (Section D)
Unomedical Devices S.A de C.V
ave.fomento industrial l9 m3
parque industrial del norte
reynosa,tamps. 88736
MX  88736
Manufacturer (Section G)
Unomedical Devices S.A de C.V
ave.fomento industrial l9 m3
parque industrial del norte
reynosa,tamps. 88736
MX   88736
Manufacturer Contact
pamela meadows
aaholmvej 1 - 3
osted
lejre 4320
DA   4320
+1(603)247-0255
MDR Report Key50317737
Report Number3003442380-2026-78207
Device Sequence Number21300539
Product Code FPA
UDI-Device Identifier05705244016682
UDI-Public05705244016682
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K032854
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 08/05/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Model Number1001682
Device Lot Number6008099
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date08/19/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured07/08/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient Age19 YR
Patient SexFemale
-
-