The customer reported that a cra-sgp, thermogard dual dispersive electrode,contact quality monitor, device was used during a cervical coolief procedure on an unknown date when it was reported, ¿patient sustained a burn beneath the grounding pad.They state the grounding pad started on the left lateral thigh of the patient, but they noticed that the grounding pad slid out of position, so they removed it and wiped the skin dry then placed in the original position.Once the procedure was finished, they noticed slight redness/warmth on the skin where the grounding pad was placed.She reports that they checked the skin again after they removed him from the room and noticed that there was some peeling up of the skin with discoloration where the pad was placed.¿ the patient is reported as ¿ok." when asked if there was medical intervention the reporter stated, ¿treated and dressed." there was no report of hospitalization for the patient.This report is being raised due to the reported injury of unknown degree of burn.
|