• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Datascope Corp. MEGA 8Fr. 50cc IAB with Accessories, US Only; System, Balloon, Intra-Aortic And Control

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Datascope Corp. MEGA 8Fr. 50cc IAB with Accessories, US Only; System, Balloon, Intra-Aortic And Control Back to Search Results
Model Number 0684-00-0296-01U
Medical Device Problem Codes Fluid/Blood Leak (1250); Difficult to Insert (1316); Unraveled Material (1664); Material Perforation (2205)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/24/2026
Type of Reportable Event Malfunction
Event or Problem Description
N/a.
 
Additional Manufacturer Narrative
Updated fields b4, d9 device available for evaluation and date return to manufacture, g3, g4 (pma 510k), g6, g7, h2, h3, h6 type of investigation, investigation findings, investigation conclusions, h11 corrected field d7a   revert a1, d4 serial number, d8 should left blank   the iab was returned with the membrane completely unfolded and small traces of blood on the exterior of the catheter.The inner lumen was found occluded with dried blood and was able to be cleared.The sheath was not returned for evaluation.  a laboratory insertion test was unable to be performed due to the membrane being unfurled.  an underwater leak test of the balloon, catheter, y fitting and extracorporeal tubing was performed and no leaks were detected.A leak may impact the ability to maintain vacuum.  the trend review did not identify a trigger of the trend data within the review period.The complaint history review did not identify an adverse trend increase in number of complaints over past three 3 months.  a dhr review was not conducted for this investigation.This complaint is not classified as a death or serious injury, nor did the investigation confirm a manufacturing defect.  we are unable to confirm the reported difficulty during insertion because of the returned condition of the catheter and we are unable to mimic the clinical setting.Per ifu the one way valve must be aspirated to 30cc while leaving the one way valve in place attached to the extracorporeal tubing luer to ensure vacuum is maintained throughout insertion.Additionally, no leaks were found which a leak could have indicated damage to the balloon.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MEGA 8Fr. 50cc IAB with Accessories, US Only
Common Device Name
System, Balloon, Intra-Aortic And Control
Manufacturer (Section D)
Datascope Corp.
15 law drive
fairfield NJ 07004
Manufacturer (Section G)
Datascope Corp.
15 law drive
fairfield NJ 07004
Manufacturer Contact
ewelina chorlowska
15 law drive
fairfield, NJ 07004
MDR Report Key50320527
Report Number2248146-2026-0001418
Device Sequence Number21236287
Product Code DSP
UDI-Device Identifier10607567109619
UDI-Public(01)10607567109619(10)3000511513(17)281107(11)251107
Combination Product (Y/N)N
Initial Reporter StateOR
Initial Reporter CountryUS
PMA/510(K) Number
K133074
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 09/30/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number0684-00-0296-01U
Device Catalogue Number0684-00-0296-01U
Device Lot Number3000511513
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer08/06/2026
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer Not provided
Supplement Date Received by Manufacturer09/29/2026
Initial Report FDA Received Date08/19/2026
Supplement Report FDA Received Date09/30/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/07/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient Age64 YR
Patient SexMale
Patient Weight221 LB
-
-