This report has been identified as b.Braun medical internal report number (b)(4).The clamp clip was unclamped and white external stopper was removed from the sample; no solution flow was observed.Crystallization was observed inside the filter and microbore tube connection.The sample was filled with red dye solution and no damage or leakage was observed at the valve and filling port, silicone sleeve and no leakage was observed between the step down tube and triangle tube.The sample was further investigated to determine any leakage point.No leakages at the connection between triangle tube and filter nor between the microbore tube and patient connector were observed.A review of the batch manufacturing record (bmr) was performed for the reported lot number, and no abnormalities or non-conformances were noted during the in process or final product inspection.Based on the investigation, the reported issue of leakage could not be observed or replicated.The complaint is considered not confirmed.We will maintain this report for further references and continue to monitor other reports for similar occurrences.If any additional pertinent information becomes available, a follow up will be submitted.
|