| Model Number |
D121 |
| Medical Device Problem Codes |
Signal Artifact/Noise (1036); Over-Sensing (1438); Pacing Problem (1439); Inappropriate/Inadequate Shock/Stimulation (1574); Connection Problem (2900)
|
| Health Effect - Clinical Code |
Shock from Patient Lead(s) (3162)
|
| Date of Event |
07/31/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that this implantable cardioverter defibrillator (icd) and the right ventricular (rv) lead delivered inappropriate anti-tachycardia pacing (atp) and a single shock due to oversensing of noise.The patient had a strong underlying rhythm, and no pacing inhibition was noted.The field representative (rep) didn't know what the patient was doing when the noise was recorded.Technical services (ts) discussed troubleshooting options to reproduce the noise and recommended asking what the patient was doing at the time of the shock.This icd remains in service.No adverse patient effects were reported.Additional information was received from the field.The patient was seen in clinic and the physician suspected subclavian crush.This icd was explanted and replaced.The rv lead was surgically abandoned and replaced.No additional adverse patient effects were reported.Addinal information was received.This icd was explanted because the header was not compatible with the new rv lead.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information added to b5.F10 and h6 updated.This product investigation is still in progress.This report will be updated when product investigation is complete.
|
| |
|
Search Alerts/Recalls
|