| Model Number |
BSM-1733 |
| Medical Device Problem Codes |
Unable to Obtain Readings (1516); Incorrect, Inadequate or Imprecise Result or Readings (1535); Output Problem (3005)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/29/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The biomedical engineer (biomed) reported that during setup the nibp pump and spo2 readings on the bsm-1733a were intermittently failing.No patient harm was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Details of complaint:
the biomedical engineer (biomed) reported that during setup the nibp pump and spo2 readings on the bsm-1733a were intermittently failing.No patient harm was reported.Investigation summary:
the complaint device was returned to nk for evaluation and exchange.During the evaluation, the reported issue could not be duplicated.As such, this complaint could not be confirmed, and the root cause of the reported issue could not be determined.During the evaluation, ecg failure was observed.The root cause of the observed ecg failure is due to damage to the ecg port.Nk will continue to monitor and trend.Additional information:
b4 date of this report
g3 date received by manufacturer
g6 type of report
h2 if follow-up, what type?
h3 device evaluated by manufacturer?
h6 event problem and evaluation codes
h11 additional manufacturer narrative.
|
| |
|