It was reported that a patient underwent a left sided nonischemic ventricular tachycardia ablation with a thermocool smarttouch catheter and the patient experienced a cardiac tamponade that required pericardiocentesis.The patient experienced a cardiac tamponade during the transseptal puncture.The cardiac tamponade stabilized by the physicians with a puncture.The procedure was successfully completed.It was reported that the thermocool smarttouch catheter was not involved according to the physicians.Additional information was received on 21-jul-2026 which indicated that the adverse event was discovered post use of biosense webster product.Seven radiofrequency ablations were performed outside the pulmonary veins.No ablations were performed within the pulmonary veins.A brk needle was used for the transseptal puncture.The flow setting was set to 0-30w 17ml/min, 30+w 30ml/min, low flow 2ml/min.After, received additional information on 21-jul-2026 which indicated that the sheath used for transseptal puncture was an agilis sheath.Ablation was performed prior to noting the cardiac tamponade.The adverse event was originally considered non-reportable under the thermocool smarttouch catheter as it was reported that the thermocool smarttouch catheter was not involved and that the event occurred during transseptal.However, bwi became aware on 21-jul-2026 that ablation occurred prior to the cardiac tamponade and therefore, the adverse event was reassessed as reportable under the thermocool smarttouch catheter.The awareness date for this record is 21-jul-2026.
|
|
G4.Pma/ 510(k): p030031;p040036.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by biosense webster, inc., or its employees that the report constitutes an admission that the product, biosense webster, inc., or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Manufacturer's reference number: (b)(4).
|