Evaluation determined that the product did not meet specifications; this is addressed by capa 12866756.Visual: received 1 all silicone foley catheter with syringe attached to a drainage bag.Verified material number 119314 and manufacturing lot number ngkx5300.Visual inspection noted the catheter balloon had ruptured.Further inspection noted the balloon had no missing pieces.Functional: inflated catheter with 10ml mbs using slip tip syringe.Upon inflation leakage at balloon due to rupture.Per capa 12866756 the potential root cause is:
silicone balloon molding pin configuration when removing the balloon from the mold.Inspection method: the inspection method is visual and relies on the operator¿s judgment to identify any leakage in the catheter balloon during the inflation/deflation inspection test at 100 percent final inspection.The labeling is found to be adequate.Dhr review is not required as event has already been investigated per capa 12866756.Corrections: d, f, h.Udi format updated with available product information per gudid.H11: sections a through f.The information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|