| Model Number |
72404251 |
| Medical Device Problem Codes |
Inflation Problem (1310); Material Puncture/Hole (1504)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/22/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that this patient's inflatable penile prothesis (ipp) cylinders failed to inflate.As a result, the device was explanted and a new ipp was implanted.After the extraction, it was observed that both cylinders showed ruptured.No further patient complications were reported.
|
| |
|
Event or Problem Description
|
|
It was reported that this patient's inflatable penile prothesis (ipp) cylinders failed to inflate.As a result, the device was explanted and a new ipp was implanted.After the extraction, it was observed that both cylinders showed ruptured.No further patient complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Model number/catalog number: 72404155.Serial number: n/a.Batch/lot number: 0167555006.Model/catalog description: reservoir 65 ml pc/iz.Unique identifier (udi) #: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
The device is not available for analysis; therefore, no physical analysis of the product could be performed.The reported device performance allegation cannot be confirmed.A risk review was completed using (b)(4) ams700 hazard analysis and confirmed that the events of "excess force or friction on silicone leads to stress, strains, holes or tears" and "mechanical malfunction (leaks, incomplete inflation/ deflation, kinking)" were defined in the product risk documentation.Based on the information available the cause that contributed to the reported event cannot be established.Therefore, a conclusion code of cause not established was assigned to this investigation.Model number/catalog number: 72404155.Serial number: n/a.Batch/lot number: 0167555006.Model/catalog description: reservoir 65 ml pc/iz.Unique identifier (udi) #: (b)(4).
|
| |
|
Search Alerts/Recalls
|