U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION AMS 700 LGX MS Pump; prosthesis, penis, inflatable

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION AMS 700 LGX MS Pump; prosthesis, penis, inflatable Back to Search Results
Model Number 72404251
Medical Device Problem Codes Inflation Problem (1310); Material Puncture/Hole (1504)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/22/2026
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that this patient's inflatable penile prothesis (ipp) cylinders failed to inflate.As a result, the device was explanted and a new ipp was implanted.After the extraction, it was observed that both cylinders showed ruptured.No further patient complications were reported.
 
Event or Problem Description
It was reported that this patient's inflatable penile prothesis (ipp) cylinders failed to inflate.As a result, the device was explanted and a new ipp was implanted.After the extraction, it was observed that both cylinders showed ruptured.No further patient complications were reported.
 
Additional Manufacturer Narrative
Model number/catalog number: 72404155.Serial number: n/a.Batch/lot number: 0167555006.Model/catalog description: reservoir 65 ml pc/iz.Unique identifier (udi) #: (b)(4).
 
Additional Manufacturer Narrative
The device is not available for analysis; therefore, no physical analysis of the product could be performed.The reported device performance allegation cannot be confirmed.A risk review was completed using (b)(4) ams700 hazard analysis and confirmed that the events of "excess force or friction on silicone leads to stress, strains, holes or tears" and "mechanical malfunction (leaks, incomplete inflation/ deflation, kinking)" were defined in the product risk documentation.Based on the information available the cause that contributed to the reported event cannot be established.Therefore, a conclusion code of cause not established was assigned to this investigation.Model number/catalog number: 72404155.Serial number: n/a.Batch/lot number: 0167555006.Model/catalog description: reservoir 65 ml pc/iz.Unique identifier (udi) #: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMS 700 LGX MS Pump
Common Device Name
prosthesis, penis, inflatable
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
Boston Scientific Corporation
300 boston scientific way
marlborough MA 01752
Manufacturer Contact
farshad fahimi
4100 hamline avenue north
building c
saint paul, MN 55112
MDR Report Key50329751
Report Number2124215-2026-44658
Device Sequence Number5508057
Product Code JCW
UDI-Device Identifier00878953003597
UDI-Public(01)00878953003597(17)190323(10)169923001
Combination Product (Y/N)Y
Initial Reporter StateNV
Initial Reporter CountryUS
PMA/510(K) Number
N970012
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/25/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Expiration Date03/23/2019
Device Model Number72404251
Device Catalogue Number72404251
Device Lot Number0169923001
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer Not provided
Supplement Date Received by Manufacturer09/03/2026
Initial Report FDA Received Date08/19/2026
Supplement Report FDA Received Date09/25/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured03/23/2017
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient Age52 YR
Patient SexMale
-
-