A lead extraction procedure commenced to remove a right ventricular (rv), right atrial (ra), and left ventricular (lv) lead due to cied system/pocket infection.During preparation of the leads, the lv lead was extracted with manual traction.Spectranetics lld ez lead locking devices (lld ezs) were inserted into the ra and rv lead to provide traction.Utilizing multiple spectranetics devices (14f glidelight laser sheath, 16f glidelight laser sheath, medium visisheath dilator sheath, large visisheath dilator sheath, 11f tightrail sub-c rotating dilator sheath, and 13f tightrail sub-c rotating dilator sheath) on the ra and rv lead, the devices were unable to pass the clavicle and first rib junction.After several attempts, both the leads broke and the decision was made to snare the leads from the groin.The ra lead was removed; however, the rv lead would not release from the clavicle/rib junction or the distal tip of the rv apex.The rv lead stretched, and broke at the clavicle, along with the lld ez.The rv lead and lld ez remnant were abandoned and remained in the patient.The patient survived the procedure.This report captures a portion of the lld ez within the rv lead that separated, along with the lead, and retained in the patient.
|
|
A3b) patient gender - not available from facility.H3/h6) a portion of the device was retained at the facility and a portion remained within the patient, thus no product investigation was performed, and the cause of the reported failure could not be established.Submission of this report does not, in itself, represent a conclusion by the manufacturer and/or authorized representative or the national competent authority that the content of this report is complete or accurate, that the medical device(s) listed failed in any manner and/or that the medical device(s) caused or contributed to an alleged death or deterioration in the state of the health of any person.
|