| Model Number |
VICM5_13.2 (-7.50) |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Intraocular Pressure Increased (1937); Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
|
| Date of Event |
05/03/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
A facility representative reported that following an implantable collamer lens (icl) implantation procedure, the patient experienced excessive vault, elevated iop, significant reduction of iridocorneal angles and significant shallowing of the ac.Medication was prescribed.Reportedly, 'iop was reduced in office same day and never elevated again' and 'pupil is fully reactive'.The lens remains implanted.The cause of the event was reported as unknown.Attempts to obtain additional information have not been successful.If additional information is received a supplemental medwatch report will be submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
H11- manufacturer narrative: iop elevation from baseline is identified in the labeling as a known potential adverse event following icl implantation.H6.Health effect- clinical code (e): 4581- 'significant reduction of iridocorneal angles'; 'shallowing of the ac'; 'reactive pupil'.H6- investigation type code (b): 4110- lens work order search-no similar complaint event(s) within associated lots were found.Claim# (b)(4).
|
| |
|