MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION AMS 700 CX MS Pump; prosthesis, penis, inflatable

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION AMS 700 CX MS Pump; prosthesis, penis, inflatable Back to Search Results
Model Number 72404233
Medical Device Problem Codes Fluid/Blood Leak (1250); Mechanical Problem (1384); Material Puncture/Hole (1504)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/26/2026
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the patient's inflatable penile prosthesis (ipp) was not functioning properly.A surgical procedure was performed, during which fluid loss was identified due to a subtotal tear in the tubing connecting the pump to the right corpus cavernosum.The only remaining connection was through the coil embedded within the tubing.The pump was found to be correctly positioned within the scrotum.All other tubing was inspected and showed no issues related to positioning or material integrity.The pump was removed and replaced, with a new connection established using the replacement pump.No patient complications were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMS 700 CX MS Pump
Common Device Name
prosthesis, penis, inflatable
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
Boston Scientific Corporation
300 boston scientific way
marlborough MA 01752
Manufacturer Contact
farshad fahimi
4100 hamline avenue north
building c
saint paul, MN 55112
MDR Report Key50331324
Report Number2124215-2026-44701
Device Sequence Number5483307
Product Code JCW
UDI-Device Identifier00878953003467
UDI-Public(01)00878953003467(17)260108(10)1100424807
Combination Product (Y/N)Y
Initial Reporter CountryGM
PMA/510(K) Number
N970012
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2025
Device Explanted Year2026
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 08/19/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Expiration Date01/08/2026
Device Model Number72404233
Device Catalogue Number72404233
Device Lot Number1100424807
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date08/19/2026
Date Device Manufactured01/09/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient SexUnknown
-
-