It was reported that during a two-level roi-c procedure, the hook of the cage holder fractured.After the blade was inserted into the second level, the hook fractured while the surgeon was withdrawing the holder from the cage.The broken hook tip was ultimately found outside of the surgical site and there was no patient or procedural impact.
|
|
Corrections in d4: lot & udi numbers, d9, and h3.Additional information in h4 and h6: investigation type, findings, and conclusions.The dhr was reviewed and there are no indications of manufacturing issues which would have contributed to this event.The device was likely conforming when it left highridge¿s control.This event likely cannot be attributed to manufacturing or design related issues.An inspection of the part revealed the hook has fractured.The complaint is confirmed for roi-c implant holder for the failure of breaks/fractures.The failure could possibly be attributed to excessive and repeated lateral movements being applied beyond intended use.This device is used for treatment.If additional information is obtained that adds value to the relevant content of this report, a follow-up report will be sent.
|