| Model Number |
CMS15-10C-US |
| Medical Device Problem Code |
Failure to Advance (2524)
|
| Health Effect - Clinical Code |
Stroke/CVA (1770)
|
| Date of Event |
08/10/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported via poster presentation that difficulty/failure to place device and stroke occurred.This presentation discussed evaluation of cases with difficulty deploying the sentinel during transfemoral transcatheter aortic valve implant procedures.A total of 102 cases in which the sentinel was used between (b)(6) 2025 and (b)(6) 2026 were analyzed to examine the anatomical characteristics and other factors in cases where placement was difficult.There were 12 cases of placement difficulty; in 5 of these, the entire device had to be removed due to the difficulty, while in 7 cases, only the proximal filter was partially placed because the distal filter could not be placed.Analysis of preoperative ct scans revealed that cases with placement difficulty involved tortuosity or calcification of the brachiocephalic artery.Cerebral infarction was observed in 2 cases within 30 days postoperatively; however, no other complications (such as vascular complications or major bleeding) were observed.
|
| |
|
Event or Problem Description
|
|
It was reported via poster presentation that difficulty/failure to place device and stroke occurred.This presentation discussed evaluation of cases with difficulty deploying the sentinel during transfemoral transcatheter aortic valve implant procedures.A total of 102 cases in which the sentinel was used between february 2025 and march 2026 were analyzed to examine the anatomical characteristics and other factors in cases where placement was difficult.There were 12 cases of placement difficulty; in 5 of these, the entire device had to be removed due to the difficulty, while in 7 cases, only the proximal filter was partially placed because the distal filter could not be placed.Analysis of preoperative ct scans revealed that cases with placement difficulty involved tortuosity or calcification of the brachiocephalic artery.Cerebral infarction was observed in 2 cases within 30 days postoperatively; however, no other complications (such as vascular complications or major bleeding) were observed.
|
| |
|
Additional Manufacturer Narrative
|
|
B3: bsc aware date used as event date is unknown.Jtvt2026 poster presentation information:
presentation number: p11-5
title: cerebral embolic protection device use in tf-tavi at the facility: evaluation of cases with difficult device deployment.With all the available information, boston scientific (bsc) concludes:
device history record review
the batch number for the sentinel cerebral protection system (csp) was not provided.A ship history was performed to identify the three most recently shipped batches to the customer.A review was completed of the device history records (dhr) for the sentinel cps batch # 39316726, 39526574, 39545253.All devices from these batches were processed through all normal operation conditions and passed all acceptance activities.The dhr reviews did not identify any deviations or non-conformances that could contribute to the event.Device technical analysis
the sentinel cps was not returned therefore analysis could not be completed.Investigation conclusion
bsc has assigned an investigation conclusion code of adverse event related to procedure.
|
| |
|
Additional Manufacturer Narrative
|
|
B3: bsc aware date used as event date is unknown.Jtvt2026 poster presentation information:
presentation number: p11-5.Title: cerebral embolic protection device use in tf-tavi at the facility: evaluation of cases with difficult device deployment.
|
| |
|