D151 |
| Medical Device Problem Codes |
Signal Artifact/Noise (1036); Over-Sensing (1438); High Capture Threshold (3266)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/20/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that electrogram (egm) review of this implantable cardioverter defibrillator (icd) system revealed possible electromagnetic interference (emi) noise with oversensing and elevated threshold measurements on the right ventricular (rv) defibrillation lead.There was no evidence of pacing inhibition or asystole, as intrinsic conduction was observed through the noise.Shock impedance trends showed a gradual increase in measurements from 50 to 70 ohms over the last seven years, consistent with calcification.This icd remains in service.No adverse patient effects were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The local area field representative was contacted for additional information regarding event cause and resolution.At this time, no further information is available.Should additional information become available this report will be updated.The product was not returned for analysis; therefore, a technical analysis could not be performed.Based on a thorough review of the reported complaint, boston scientific has assigned an investigation conclusion code of known inherent risk of device.
|
| |
|