• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MED-EL Elektromedizinische Geraete GmbH SYNCHRONY; Cochlear Implant

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MED-EL Elektromedizinische Geraete GmbH SYNCHRONY; Cochlear Implant Back to Search Results
Model Number Mi1250 SYNCHRONY 2
Medical Device Problem Code Device Appears to Trigger Rejection (1524)
Health Effect - Clinical Code Tissue Breakdown (2681)
Date of Event 08/10/2026
Type of Reportable Event Serious Injury
Event or Problem Description
Device has extruded through the skin flap.There was also an infection.Additionally, there was a partial migration of the electrode out of cochlea.Hearing performance with the device is also affected.
 
Event or Problem Description
Device has extruded through the skin flap.There was also an infection.Additionally, there was a partial migration of the electrode out of cochlea.Hearing performance with the device is also affected.Explantation is planned.
 
Additional Manufacturer Narrative
The device has not been explanted.If it should be explanted, it should be returned to the manufacturer for evaluation.When available, a device failure analysis will be submitted as a follow up report.
 
Additional Manufacturer Narrative
Additional information: according to information received from from the field, the device has extruded through the skin flap and an infection was observed.A review of the device_s sterilization records shows that the device has been subject to a valid sterilization process, thus no available information points to the implant being the source of the reported issue.Therefore it is unlikely that the device was the source of the reported infection.However, a device contribution to the subsequent development of the extrusion cannot be ruled out.In addition a partial migration of the electrode out of the cochlea is reported.Device explantation is planned.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNCHRONY
Common Device Name
Cochlear Implant
Manufacturer (Section D)
MED-EL Elektromedizinische Geraete GmbH
innsbruck
AU 
Manufacturer Contact
laura simonotti
fuerstenweg 77a
innsbruck 6020
AU   6020
+43()577885626
MDR Report Key50338450
Report Number9710014-2026-00769
Device Sequence Number5725798
Product Code MCM
UDI-Device Identifier09008737395452
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
P000025
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 09/24/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date08/10/2025
Device Model NumberMi1250 SYNCHRONY 2
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer Not provided
Supplement Date Received by Manufacturer08/14/2026
Initial Report FDA Received Date08/20/2026
Supplement Report FDA Received Date09/24/2026
Date Device Manufactured08/10/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age58 YR
Patient SexMale
-
-