| Catalog Number |
MC1816 |
| Medical Device Problem Code |
Failure to Fire (2610)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/30/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
On (b)(6), 2026, a patient underwent an ultrasound guided kidney biopsy procedure through tissue of normal density using a maxcore instrument.During the procedure, the device cannula failed to fire, and the firing button became stuck, although it could still be pressed.No coaxial needle was used.The procedure was completed using another device.There was no reported patient injury.This is report 6 of 7.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2026, a patient underwent an ultrasound guided kidney biopsy procedure through tissue of normal density using a maxcore instrument.During the procedure, the device outer cannula failed to fire, and the firing bottom button became stuck, although it could still be pressed.No coaxial needle was used.The procedure was completed using another device.There was no reported patient injury.This is report 6 of 7.
|
| |
|
Additional Manufacturer Narrative
|
|
H11: as the lot number for the device was provided, a review of the device history records was performed.The device has not been returned to the manufacturer for evaluation.However, photos were provided for review.The investigation of the reported event is currently underway.Section a through f: the information provide by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Additional Manufacturer Narrative
|
|
H11: manufacturing review: the device history records have been reviewed, and this lot met all release criteria.Investigation summary: the physical device was not returned for evaluation.One photo was provided and reviewed.In the photo, seven maxcore devices were visible with the packaging and labelling details noted to be sitting on a flat surface.Verified the same with trackwise.All the maxcore devices were in initial position with needle protective tubing and with yellow guard attached to the devices.No other anomalies were noted.Based on the review, reported event cannot be confirmed.Therefore, the investigation is inconclusive for the reported failure to fire as no objective evidence was shown in the provided photo to support the reported event for all the devices.A definitive root cause for the alleged failure to fire could not be determined based upon the provided information.Labeling review: as the reported event did not allege a labeling or use related issue, a labeling review is not required.B5, g3, h6 (component, method, result, conclusion).Section a through f: the information provide by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Search Alerts/Recalls
|
|
|