| Model Number |
PFA-GEN-01 |
| Medical Device Problem Code |
Thermal Decomposition of Device (1071)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/01/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
This report is to advise of thermal damage found during investigation.During a procedure with tactiflex duo catheter, the catheter was successfully connected to the current generator and placed in the heart.The energy was charged to 100%.Before the first ablation, a loud sound was heard from the generator and an error message appeared which had a restart button.The exact message was not clear but it mentioned "low voltage detected".The generator was restarted two times with no resolution and another error appeared saying "hardware issue detected".The generator was replaced and the procedure was completed with no adverse consequences to the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
One current pfa generator was received for evaluation.Based on the information provided to abbott and the investigation performed, the root cause can be attributed to thermal damage to the pinnacle bridge pca and pinnacle bfc pca.The batch record was reviewed to ensure that each manufacturing and inspection operation was performed.No non-conformances associated with the reported event were identified.
|
| |
|