• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Medtronic Puerto Rico Villalba CapSure Epicardial Lead; PERMANENT PACEMAKER ELECTRODE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Medtronic Puerto Rico Villalba CapSure Epicardial Lead; PERMANENT PACEMAKER ELECTRODE Back to Search Results
Model Number 4968-25
Medical Device Problem Codes Failure to Capture (1081); Fracture (1260); High impedance (1291); Under-Sensing (1661)
Health Effect - Clinical Code Dizziness (2194)
Date of Event 07/29/2026
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the patient experienced dizziness.It was further reported that the right ventricular (rv) lead exhibited no capture, no sensing, high impedance and was fractured.The lead was capped and replaced. no patient complications have been reported as a result of this event.
 
Additional Manufacturer Narrative
Continuation of d10: d121 ipg, implanted on (b)(6) 2024 medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CapSure Epicardial Lead
Common Device Name
PERMANENT PACEMAKER ELECTRODE
Manufacturer (Section D)
Medtronic Puerto Rico Villalba
road 149 km 56.3
villalba PR 00766
Manufacturer (Section G)
Medtronic Puerto Rico Villalba
road 149 km 56.3
villalba PR 00766
Manufacturer Contact
gerard torenvliet
8200 coral sea st ne
mounds view, MN 55112--439
+1(763)526-9277
MDR Report Key50344773
Report Number2649622-2026-24094
Device Sequence Number16010566
Product Code DTB
UDI-Device Identifier00613994492272
UDI-Public(01)00613994492272(17)180209
Combination Product (Y/N)N
Initial Reporter StateOR
Initial Reporter CountryUS
PMA/510(K) Number
P950024
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Device Explanted Year2026
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 08/20/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Expiration Date02/09/2018
Device Model Number4968-25
Device Catalogue Number4968-25
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Fifteen-Day
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date08/20/2026
Date Device Manufactured03/08/2016
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
0181 LEAD
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient Age15 YR
Patient SexMale
-
-