| Model Number |
IntelliVue Patient Monitor MX500 |
| Medical Device Problem Code |
Application Program Freezes or Fails to Launch (4031)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/27/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the monitor was completely frozen.It was unknown whether the parameters and waveforms were also frozen.The device was in use on a patient at the time of the event.There was no adverse event reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The functional test was carried out by the customer.The rse confirmed that the reported issue is related to a problem in the intellivue patient monitor software which has been fixed in software patch r.00.03.Based on the results of the analysis, it was determined that the product has malfunctioned and the most likely cause of the reported problem was a software problem.The issue was resolved by updating the monitor to software revision r.00.03.Reporting institution phone#: (b)(6).
|
| |
|