Model Number 0684-00-0576-01U
Medical Device Problem Codes Signal Artifact/Noise (1036); Break (1069); Obstruction of Flow (2423); Material Deformation (2976)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/17/2026
Type of Reportable Event Malfunction
Event or Problem Description
N/a.
 
Additional Manufacturer Narrative
Updated fields: b4, b7, d10, g3, g6, g7, h2, h3, h6 (type of investigation, investigation findings, investigation conclusions, component code), h11.Product was not returned so an exact cause of the issue could not be identified.However, based on our investigation, a unable to update timing issue can occur due to one or more of the following probable causes.Component failure - potential root cause breakdown: pressure signal unavailable or unreliable.Primary root cause category: timing cannot be updated because the arterial waveform required for timing reference is incorrect or absent.Fiber-optic sensor-related causes: root cause: fiber-optic arterial signal is not detected or is invalid.Contributing factors: fiber-optic calibration failure.Fiber-optic sensor disconnection.Fiber-optic cable break.Fiber-optic cable kink or excessive bending.Improper connector insertion or unsecured cable.Contamination or damage to exposed sensor tip.Use of fiber-optic iab catheter on non-maquet-compatible pumps (off-label).Inner lumen-related causes: root cause: inner lumen cannot transmit an accurate arterial pressure waveform.Contributing factors: blood clot (thrombus) formation within the inner lumen.Partial or complete inner lumen occlusion.Kinked inner lumen.Break or damage to the inner lumen wall.Resistance encountered during aspiration (indicative of occlusion).Using the inner lumen for blood sampling (degrades waveform quality).Inadequate continuous flush (not maintained at 3 cc/hr).Failure to aspirate and discard 3 cc prior to setup.Air bubbles introduced during setup or flushing.Failure to stop pumping during fast flush (embolism prevention step ignored).User error - potential root cause breakdown: pressure signal unavailable or unreliable.Primary root cause category.Timing cannot be updated because the arterial waveform required for timing reference is incorrect or absent.Operation mode/user interaction factors: root cause: timing adjustments cannot be performed correctly due to signal or workflow issues.Contributing factors.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Sensation Plus 8Fr. 50cc IAB with Accessories, US Only
Common Device Name
System, Balloon, Intra-Aortic And Control
Manufacturer (Section D)
Datascope Corp.
15 law drive
fairfield NJ 07004
Manufacturer (Section G)
Datascope Corp.
15 law drive
fairfield NJ 07004
Manufacturer Contact
ewelina chorlowska
15 law drive
fairfield, NJ 07004
MDR Report Key50346981
Report Number2248146-2026-0001427
Device Sequence Number17652125
Product Code DSP
UDI-Device Identifier10607567109633
UDI-Public(01)10607567109633(10)3000523482(17)290112(11)260112
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
K133074
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 09/24/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number0684-00-0576-01U
Device Catalogue Number0684-00-0576-01U
Device Lot Number3000523482
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer08/24/2026
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer Not provided
Supplement Date Received by Manufacturer09/01/2026
09/22/2026
Initial Report FDA Received Date08/20/2026
Supplement Report FDA Received Date09/21/2026
09/24/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/12/2026
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
Intra-Aortic Balloon Pump
Patient Age55 YR
Patient SexMale
-
-