MAUDE Adverse Event Report: Stryker GmbH IMN Instruments - Distal Targeting Arm Femur Antegrade; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Stryker GmbH IMN Instruments - Distal Targeting Arm Femur Antegrade; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES Back to Search Results
Catalog Number 23533106
Medical Device Problem Code Failure to Align (2522)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/23/2026
Type of Reportable Event Malfunction
Event or Problem Description
As reported: "during insertion of a distal locking screw using the gamma4 distal targeting device, drilling was performed with a locking drill.While advancing through the medial cortex, an audible sound was noted, and fluoroscopy confirmed that the drill had deviated posterior to the nail.The drill was withdrawn and drilling was reattempted.During the second attempt, a loud sound was heard and the drill stopped within the nail.Fluoroscopic evaluation confirmed that the drill remained within the intended trajectory, and drilling was continued.A subsequent loud sound was heard, and fluoroscopy revealed fracture of the drill approximately 25 mm from the tip while inside the nail.The fractured fragment was retrieved using forceps and a k-wire.The procedure was completed successfully without further complications.".
 
Additional Manufacturer Narrative
Once the investigation has been completed any additional information will be reported in a supplemental report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMN Instruments - Distal Targeting Arm Femur Antegrade
Common Device Name
ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES
Manufacturer (Section D)
Stryker GmbH
bohnackerweg 1
postfach
selzach 2545
SZ  2545
Manufacturer (Section G)
Stryker Trauma GmbH
prof. kuentscher-strasse 1-5
schoenkirchen/kiel D-242 32
GM   D-24232
Manufacturer Contact
dannielle stap
325 corporate drive
mahwah, NJ 07430
+1(201)831-5000
MDR Report Key50348319
Report Number0008031020-2026-01185
Device Sequence Number20343406
Product Code HSB
UDI-Device Identifier07613327357066
UDI-Public07613327357066
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K172774
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 08/20/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number23533106
Device Lot NumberKME921957
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date08/20/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient Age70 YR
Patient SexMale
-
-