| Model Number |
40495 |
| Medical Device Problem Code |
Material Integrity Problem (2978)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/22/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The coating of the wire distal is removed from the patient's body during the procedure.The 70% stenosed target lesion was located in a moderately tortuous and moderately calcified vessel.A 300cm, 8cm v-18 control wire was selected for use in an angioplasty procedure in the leg.During the procedure, the distal coating of the guidewire was removed from the patients body.The procedure was completed using with another of the same device and there were no patient complications as a result of this event.It was further reported that the distal coating of the guidewire appeared to peel.Resistance or unusual difficulty was encountered during advancement, manipulation of the guidewire.The coating removal occurred at the distal tip of the guidewire.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information provided in the b5.Investigation results: device history records (dhr) review: it was confirmed this device met manufacturing specifications prior to distribution and there were no manufacturing deviations that could have contributed to the reported event.Investigation conclusion: based on a thorough review of the reported complaint, the most probable cause for this complaint was cause not established.
|
| |
|