Zoll medical corporation evaluated the device and the device performed to specification.The customer's reports were not replicated or confirmed.The logs showed that the automatic readiness test began failing after the pads were changed from onestep complete pads to onestep basic pediatric pads.Although the onestep basic pediatric pads may be used for manual testing, they were not compatible with the device's current configuration for completing the automatic readiness test.Based on the evaluation the report of failed self test is closed as device performed as expected.Review of the device activity logs did not show any occurrences or evidence to support the customer's report of pacer disables faults.Therefore, our investigation for the reported malfunction of pacer disabled was unsubstantiated.The device was recertified and returned to the customer.No trend is associated with reports of these types.
|