| Model Number |
0295 |
| Medical Device Problem Code |
Impedance Problem (2950)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/24/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that this right ventricular (rv) lead was surgically abandoned during a device replacement procedure.The device was replaced due to reaching elective replacement indicator (eri).The physician chose to implant a new rv lead instead of using this rv lead due to inconsistent impedance measurements.The physician suspected that trying to remove or using this rv lead with the new implant would cause damage.This rv lead was successfully replaced.No additional adverse patient effects were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was surgically abandoned; therefore, a technical analysis could not be performed.Based on a thorough review of the reported complaint, boston scientific has assigned an investigation conclusion code of known inherent risk of device.
|
| |
|