• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Curonix LLC Freedom Peripheral Nerve Stimulator

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Curonix LLC Freedom Peripheral Nerve Stimulator Back to Search Results
Model Number STQ4-RCV-A0
Medical Device Problem Code Use of Device Problem (1670)
Health Effect - Clinical Code Shock from Patient Lead(s) (3162)
Date of Event 08/05/2026
Type of Reportable Event Serious Injury
Event or Problem Description
The patient reported periodic uncomfortable stimulation.The frequency was lowered on the transmitter assemblies and as of (b)(6) 2026, the shocking sensations have gone away and the patient is getting marginal relief.
 
Additional Manufacturer Narrative
The unintended stimulation/new pain questionnaire was reviewed for potential causes of the reported issue.Based on this review, the patient having another non-curonix device implanted was ruled out.However, after lowering the frequency on the transmitter assemblies, the shocking sensations have gone away and the patient is getting marginal relief.The stimulator is used to treat pain.The cause of the reported issue is due to programming parameters as the shocking sensations were resolved after lowering the frequency on the transmitter assemblies (user error-commercial team member).Rates reviewed at the most recent complaint trending meeting do not indicate a significant increase in unintended stimulation/new pain issues.Unintended stimulation/new pain issue rates remain acceptably low; thus, a capa is not required.Unintended stimulation/new pain issues rates will continue to be tracked and trended.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Freedom Peripheral Nerve Stimulator
Common Device Name
Peripheral Nerve Stimulator
Manufacturer (Section D)
Curonix LLC
1310 park central boulevard s.
pompano beach FL 33064
Manufacturer (Section G)
Curonix LLC
1310 park central boulevard s.
pompano beach FL 33064
Manufacturer Contact
alicia pagliaro
1310 park central boulevard s.
pompano beach, FL 33064
+1(800)965-5134
MDR Report Key50356236
Report Number3010676138-2026-00172
Device Sequence Number21360213
Product Code GZF
UDI-Device Identifier00850051034406
UDI-Public(01)00850051034406(17)280201
Combination Product (Y/N)N
Initial Reporter StateWV
Initial Reporter CountryUS
PMA/510(K) Number
K233162
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2026
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 08/20/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Model NumberSTQ4-RCV-A0
Device Lot NumberSWO260223
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date08/20/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured02/02/2026
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
Transmitter Assembly SN:(b)(6) LN: 01704-DT; Transmitter Assembly SN:(b)(6) LN: 01704-DT
Patient Age74 YR
Patient SexMale
-
-