| Model Number |
3386 |
| Medical Device Problem Code |
High impedance (1291)
|
| Health Effect - Clinical Code |
Inadequate Pain Relief (2388)
|
| Date of Event |
07/23/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that following a fall the patient experienced ineffective therapy.Diagnostics revealed high impedances on the lead.Reprogramming has been unable to resolve the issue.As a result, surgical intervention was undertaken on (b)(6) 2026 wherein the lead and both extensions were explanted and replaced to address the issue.
|
| |
|
Additional Manufacturer Narrative
|
|
Date of event is estimated.The results of the investigation are inconclusive since the devices were not returned for analysis.Based on the information received, the cause of the reported event could not be conclusively determined.
|
| |
|