Abbott Laboratories Alinity c Processing Module; ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE
|
Back to Search Results |
|
| Catalog Number |
03R67-01 |
| Medical Device Problem Code |
Low Test Results (2458)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/10/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer observed depressed sodium when processing on alinity c processing module.On (b)(6) 2026, sid (b)(6) generated sodium of 102 mmol/l.The result was reported out of lab and patient was referred to intensive care unit.No treatment was given.On (b)(6) 2026, the same patient generated sodium of 140 mmol/l, matching historical patient sodium of 140 mmol/l.The customer normal sodium range is >125 mmol/l.No impact on patient management was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The alinity c processing module was inspected, and the ict probe was replaced which resolved the issue.The module function was verified with a control run.An instrument service history review for serial number ac06925 revealed no additional discrepant sodium results reported after the current ticket was reported.A review of tracking and trending for the ict probe did not identify any trends.A review of tracking and trending for the alinity c processing module did not identify any trends related to the complaint issue.A review of manufacturing documentation did not identify any non-conformances or potential non-conformances for the ict probe.Labeling was reviewed and found to adequately address the complaint issue.Based on the investigation, no systemic issue or deficiency of the alinity c processing module serial number ac06925 or the ict probe was identified.
|
| |
|
Search Alerts/Recalls
|
|
|