• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Encore Medical L.P. DJO SURGICAL; EMPOWR 3D KNEETM INS, 9L 10MM, VE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Encore Medical L.P. DJO SURGICAL; EMPOWR 3D KNEETM INS, 9L 10MM, VE Back to Search Results
Catalog Number 341-10-709
Medical Device Problem Code No Apparent Adverse Event (3189)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 08/03/2026
Type of Reportable Event Serious Injury
Event or Problem Description
Revision surgery - due to unknown reason.
 
Additional Manufacturer Narrative
The agent reported, "called for revision age 55 gender male".The previous surgery and the surgery detailed in this event occurred 4 years, 8 months, 25 days apart.Item number: 341-10-709 "item description: empowr 3d kneetm ins, 9l 10mm, ve " lot#: 069t1090 part revision: b product type: knee manufacture date: 12-dec-2020 expiration date: 08-dec-2025.This evaluation is limited in scope as the item(s) associated with this investigation was not returned to djo surgical - austin for examination.The surgery was completed as intended.If additional information regarding the reported event is submitted at a future date, this investigation will be re-evaluated.There was no information submitted with this complaint about any patient activities, accidents, or medical contraindications that may have contributed to the reported event.There were no findings during this evaluation that indicate that the reported device(s) was defective.The surgeon performed this procedure to remedy the patient's condition.No further action is deemed necessary.Root cause: the root cause of this complaint was a revision surgery due to the replacement of the insert component.Multiple factors outside enovis surgical's control may have contributed to this event.Containment: inventory containment is not required as there are no indications of a product or process issue affecting implant safety or effectiveness.The revision surgery was completed successfully.Device history records review: a review of the device history record(s) shows that the reported component(s) used in the previous surgery met design and manufacturing requirements at the time of release for use.There were no nonconforming material reports associated with the product(s) that may have contributed to the reported event.The device(s) was verified to have gone through an acceptable sterilization process and was within its expiration date at the time of the previous surgery.Complaint history: customer complaint history of the reported device(s) showed no present trends or ongoing issues that need review.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DJO SURGICAL
Common Device Name
EMPOWR 3D KNEETM INS, 9L 10MM, VE
Manufacturer (Section D)
Encore Medical L.P.
9800 metric blvd.
austin TX 78758
Manufacturer (Section G)
Encore Medical L.P.
9800 metric blvd.
austin TX 78758
Manufacturer Contact
stacey bonnell
9800 metric blvd.
austin, TX 78758
MDR Report Key50358165
Report Number1644408-2026-01587
Device Sequence Number10112135
Product Code JWH
UDI-Device Identifier00888912166867
UDI-Public00888912166867
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K162024
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2026
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 08/20/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number341-10-709
Device Lot Number069t1090
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date08/20/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured12/12/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age55 YR
Patient SexMale
-
-