It was reported that a kink occurred, and the procedure was cancelled.A left atrial appendage (laa) closure procedure was being performed, and a watchman fxd curve access system (was) was selected to be used.The patient was sedated with general anesthesia.During the procedure, transeptal puncture (tsp) was performed in an inferior/posterior location and the physician approached using a was.An attempt was made to deploy the 40mm closure device; however, interference from the posterior pectinate caused the closure device to pop out, resulting in significant protrusion.The was sheath was severely bent and replacing it was unlikely to resolve the issue.Furthermore, due to insufficient space to select an alternative tsp site, the procedure was discontinued at the physician's discretion.It was further reported that there was no report that the watchman access system was kinked and that the report does not indicate that the 40mm closure device and delivery system was bent.It was reported that the sheath was on a severely angulated trajectory, and changing the sheath was not expected to resolve the issue.Therefore, the complaint will be considered non-reportable.
|