Model Number D022
Medical Device Problem Codes Pacing Problem (1439); Inappropriate/Inadequate Shock/Stimulation (1574)
Health Effect - Clinical Code Shock from Patient Lead(s) (3162)
Date of Event 07/24/2026
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that this implantable cardioverter defibrillator (icd) delivered potential inappropriate therapy for an atrial arrythmia.Inappropriate atp and shock therapy.Technical services (ts) reviewed the available device data and determined that the episodes were consistent with atrial fibrillation (af) with rapid ventricular response (rvr).Two episodes resulted in a total of two anti-tachycardia pacing (atp) bursts and four electric shocks delivered.Reprogramming options were provided and the physician planned to initiate amiodarone therapy.No additional adverse patient effects were reported.This icd remains in service.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DYNAGEN? MINI ICD DR
Common Device Name
Implantable Cardioverter Defibrillator (ICD)
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
Boston Scientific Corporation
cashel road
clonmel E91 T 862
EI   E91 T862
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
+1(651)582-6168
MDR Report Key50358902
Report Number2124215-2026-45012
Device Sequence Number10103231
Product Code LWS
UDI-Device Identifier00802526533907
UDI-Public(01)00802526533907(17)271015
Combination Product (Y/N)N
Initial Reporter StateVA
Initial Reporter CountryUS
PMA/510(K) Number
P960040
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2026
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 08/20/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberD022
Device Catalogue NumberD022
Device Lot Number143560
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date08/20/2026
Date Device Manufactured10/21/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age64 YR
Patient SexFemale
-
-