Model Number CD3369-40Q
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 07/18/2026
Type of Reportable Event Death
Event or Problem Description
It was reported that the patient deceased.There is no allegation from a healthcare professional that the death was device related.The cause of death was congestive heart failure, coronary artery disease, hypertension, atrial fibrillation, diabetes, chronic obstructive pulmonary disease, and natural causes.
 
Additional Manufacturer Narrative
Device was returned due to patient death.The device was received in normal range of operation.Analysis of device image was performed and no anomalies were noted.Further electrical testing was performed, and no issues were noted.Longevity assessment found the device was operating above the elective replacement indicator (eri) level and within expected longevity.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Quadra Assura MP ICD
Common Device Name
Defibrillator, Automatic Implantable Cardioverter, With Cardiac Resynchronization (Crt-D)
Manufacturer (Section D)
Abbott Medical (CRM-Sylmar)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
Abbott Medical (CRM-Sylmar)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
navpreet singh
15900 valley view court
sylmar, CA 91342
MDR Report Key50359003
Report Number2017865-2026-17572
Device Sequence Number15989088
Product Code NIK
UDI-Device Identifier05414734508377
UDI-Public(01)05414734508377(10)A000131601(17)240930
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
P030054
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Device Explanted Year2026
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 08/20/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date09/30/2024
Device Model NumberCD3369-40Q
Device Catalogue NumberCD3369-40Q
Device Lot NumberA000131601
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer07/31/2026
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date08/20/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured10/05/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Death;
Patient Age72 YR
Patient SexMale
-
-