| Model Number |
MD344 |
| Medical Device Problem Code |
Material Puncture/Hole (1504)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/20/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Dealer stated customer reported holes in filter after sterilization.
|
| |
|
Additional Manufacturer Narrative
|
|
No similar complaints have been recorded for md344, lot: 2602109.Dhr has been reviewed, no discrepancies noted.Image has been provided for review, image shows burn holes lining up with the retention plate.Potential root cause is excess heat/energy being put into the material causing the melt pattern consistent with the areas of highest exposure (vent holes in the sterilcontainer).Not a manufacturing defect.No further action will be taken at this time.
|
| |
|