| Model Number |
720185-01 |
| Medical Device Problem Code |
Inflation Problem (1310)
|
| Health Effect - Clinical Code |
Capsular Contracture (1761)
|
| Date of Event |
07/24/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that the patient's inflatable penile prosthesis (ipp) exhibited auto inflation.A surgical procedure was performed, during which a capsular formation surrounding the reservoir was identified.The reservoir was laparoscopically released from the capsule, which resolved the auto inflation issue.Additionally, the patient requested replacement of the existing ms pump with a tenacio pump.As a result, the pump was exchanged for a tenacio pump during the same procedure.No additional patient complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Concomitant product implanted at time of the event:
model number/catalog number: 72404234-14,
serial number: n/a,
batch/lot number: 1100414275,
model/catalog description: cx preconnect ms 24cm ps 14cm tl iz,
unique identifier (udi): (b)(4).
|
| |
|