• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION INCEPTA? ICD; Implantable cardioverter defibrillator (ICD)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION INCEPTA? ICD; Implantable cardioverter defibrillator (ICD) Back to Search Results
Model Number E160
Medical Device Problem Codes Pacing Problem (1439); Defibrillation/Stimulation Problem (1573); Inappropriate/Inadequate Shock/Stimulation (1574); Failure to Read Input Signal (1581)
Health Effect - Clinical Code Shock from Patient Lead(s) (3162)
Date of Event 08/05/2026
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that this implantable cardioverter defibrillator (icd) delivered five shocks and several rounds of anti-tachycardia pacing (atp) for what appeared to be atrial fibrillation (af) with rapid ventricular response (rvr).Technical services (ts) was contacted and after reviewing the data noted there was an irregular morphology present and that it appeared to become faster after a shock.This icd remains in service.No additional adverse patient effects were reported.
 
Additional Manufacturer Narrative
This product investigation is still in progress.This report will be updated when product investigation is complete.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INCEPTA? ICD
Common Device Name
Implantable cardioverter defibrillator (ICD)
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
Boston Scientific Corporation
cashel road
clonmel E91 T 862
EI   E91 T862
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
+1(651)582-6168
MDR Report Key50360078
Report Number2124215-2026-45040
Device Sequence Number12271814
Product Code LWP
UDI-Device Identifier00802526480751
UDI-Public(01)00802526480751(17)160205
Combination Product (Y/N)N
Initial Reporter StateWA
Initial Reporter CountryUS
PMA/510(K) Number
P960040
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2014
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 08/20/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date02/05/2016
Device Model NumberE160
Device Catalogue NumberE160
Device Lot Number109072
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date08/20/2026
Date Device Manufactured02/11/2014
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age82 YR
Patient SexFemale
-
-