• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION Autotome RX 44; Unit, Electrosurgical, Endoscopic (With Or Without Accessories)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION Autotome RX 44; Unit, Electrosurgical, Endoscopic (With Or Without Accessories) Back to Search Results
Model Number M00545170
Medical Device Problem Codes Break (1069); Deformation Due to Compressive Stress (2889)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/27/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported to boston scientific corporation that an autotome rx 44 was attempted to be used in the duodenal papilla during an endoscopic retrograde cholangiopancreatography (ercp) procedure to treat bile duct stones performed on (b)(6) 2026.During the procedure, the cutting wire was fractured when the direction of the bow was adjusted for the first time.A photo of the complaint device was provided and showed that the cutting wire was broken and kinked.The procedure was completed with another of the same device.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be stable.
 
Additional Manufacturer Narrative
Block e1: initial reporter facility name is (b)(6); reported here as the facility name exceeded the character limit for the designated field.Block h6: imdrf device code a0401 captures the reportable event of cutting wire break.Imdrf device code a040601 captures the reportable event of cutting wire kinked.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Autotome RX 44
Common Device Name
Unit, Electrosurgical, Endoscopic (With Or Without Accessories)
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
Boston Scientific Corporation
2546 calle primera
alajuela
CS  
Manufacturer Contact
christina lam
300 boston scientific way
marlborough, MA 01752
+1(914)571-6918
MDR Report Key50360700
Report Number3005099803-2026-03429
Device Sequence Number18706099
Product Code KNS
UDI-Device Identifier08714729444749
UDI-Public(01)08714729444749(17)281126(10)38062238
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K013153
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/30/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberM00545170
Device Catalogue NumberM00545170
Device Lot Number0038062238
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer Not provided
Supplement Date Received by Manufacturer09/07/2026
Initial Report FDA Received Date08/21/2026
Supplement Report FDA Received Date09/30/2026
Date Device Manufactured11/27/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age74 YR
Patient SexMale
Patient Weight110 LB
-
-