| Model Number |
MR8-AF01 |
| Medical Device Problem Codes |
Detachment of Device or Device Component (2907); Naturally Worn (2988)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/21/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that prior to use, the attachment had a bearing issue.There was no patient involved.
|
| |
|
Additional Manufacturer Narrative
|
|
H3: evaluation able to confirm reported fault of bad bearings.The likely cause of the failure was due to corrosion.It was also noted that the bearings were broken.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|