MAUDE Adverse Event Report: Biosense Webster Inc THERMOCOOL SMARTTOUCH SF; Cardiac Ablation Percutaneous Catheter

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Biosense Webster Inc THERMOCOOL SMARTTOUCH SF; Cardiac Ablation Percutaneous Catheter Back to Search Results
Catalog Number D134805
Medical Device Problem Code Obstruction of Flow (2423)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/27/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that a patient underwent an atrioventricular nodal reentrant tachycardia (avnrt) ablation procedure with a thermocool smarttouch sf catheter which had an irrigation issue during use on the patient.It was initially reported by the customer that the new thermocool smarttouch® sf catheter was not irrigating at all and no fluid was coming out of the tip of the thermocool smarttouch® sf catheter.The caller reported that the lab staff attempted to aspirate the thermocool smarttouch® sf catheter, without resolution.The thermocool smarttouch® sf catheter was replaced and the issue was resolved.The procedure continued with no further issues.There was no patient consequence.The customer's reported occlusion was not considered to be mdr reportable since the potential risk that it could cause or contribute to a serious injury or death to the operator or patient is remote.On 29-jul-2026, additional information was received indicating the catheter was used on the patient and suspected to have the irrigation issue.There was no errors noted from the ngen pump and when ablating, there was a high temperature error.It was then noticed that the tubing had disconnected from the catheter.The catheter was then removed from the body and flushing was attempted unsuccessfully.The correct catheter settings were set on the generator.Based on the information indicating the catheter was used on the patient, the event was reassessed as an mdr reportable malfunction.
 
Additional Manufacturer Narrative
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by biosense webster inc., or its employees that the report constitutes an admission that the product, biosense webster inc., or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Manufacturer's ref.#: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
THERMOCOOL SMARTTOUCH SF
Common Device Name
Cardiac Ablation Percutaneous Catheter
Manufacturer (Section D)
Biosense Webster Inc
31 technology drive, suite 200
irvine CA 92618
Manufacturer (Section G)
Biosense Webster Inc (Juarez)
circuito interior norte
1820parque industrial salvacar
juarez 32599
MX   32599
Manufacturer Contact
katie seppa
31 technology dr
irvine, CA 92618
+1(513)337-8052
MDR Report Key50361710
Report Number3013300026-2026-02734
Device Sequence Number16118383
Product Code LPB
UDI-Device Identifier10846835010183
UDI-Public(01)10846835010183(17)290324(10)31895164L
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
P030031
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberD134805
Device Lot Number31895164L
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date08/21/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured03/25/2026
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-