| Model Number |
500001002502 |
| Medical Device Problem Code |
Loose or Intermittent Connection (1371)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/28/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that when using the bd pyxis¿ medstation¿ es auxiliary, tower door was failed.The customer stated that the malfunction occurred when a user tried to issue medication to the patient and caused a delay in patient care.However, there was no adverse events or injuries reported based on this event.
|
| |
|
Additional Manufacturer Narrative
|
|
A review of the complaint history for sn (b)(6) was performed in salesforce which did not locate similar complaint(s) with the same failure mode for this serial number.A review of the device history record for sn (b)(6) was performed from the date of manufacture, 14-sep-2009 and confirmed that this device was not previously returned for servicing and there were no production failures which correlates to the customer reported issue.Upon investigation of the actual device used in this incident, it was determined that the door was failed.A field service engineer (fse) re-crimped and greased all latches to resolve the issue.The system functioned as intended after the field service engineer repaired the device.
|
| |
|
Search Alerts/Recalls
|