• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION OptiCross? 6 HD; transducer, ultrasonic, diagnostic

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION OptiCross? 6 HD; transducer, ultrasonic, diagnostic Back to Search Results
Model Number H74939354090
Medical Device Problem Code Entrapment of Device (1212)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/31/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that a catheter issue occurred.An opticross 6 hd imaging catheter was selected for ultrasound examination of the target lesion.During a percutaneous coronary intervention, the catheter became wrapped around the guidewire.The device was removed.There were no patient complications reported.
 
Additional Manufacturer Narrative
Detailed product information was not provided to bsc.We completed a good faith effort to obtain the information, but it was unable to be obtained.Because the product is unknown, we are unable to provide the unique identifier (udi) # and other specific product information.Good faith effort attempts were made to try and retrieve additional details regarding the reported event, but further information was unable to be obtained.Investigation results: device analysis findings: the device was not returned for analysis; therefore, a technical analysis could not be performed.Device history record (dhr) review: review confirmed that the device met all material, assembly, and performance specifications prior to distribution.Investigation conclusion: this investigation has been assigned an investigation conclusion code of cause not established following a review of the reported event details, available information, and product record review.In this case the device was not returned for product analysis therefore limiting the identification of a most probable cause of the reported event.Improper handling, device manipulation or not following the correct instructions during the procedure, alongside the anatomical conditions of the patient, could be contributing factors to the reported issue.However, there is no additional detail pertinent to the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OptiCross? 6 HD
Common Device Name
transducer, ultrasonic, diagnostic
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
Boston Scientific Corporation
2546 calle primera
alajuela
CS  
Manufacturer Contact
mark mischke
4100 hamline ave n
arden hills, MN 55112
+1(612)875-7234
MDR Report Key50370180
Report Number2124215-2026-45101
Device Sequence Number13255474
Product Code OBJ
UDI-Device Identifier00191506035457
UDI-Public(01)00191506035457
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
K213593
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 09/24/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberH74939354090
Device Catalogue NumberH74939354090
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer Not provided
Supplement Date Received by Manufacturer09/03/2026
Initial Report FDA Received Date08/21/2026
Supplement Report FDA Received Date09/24/2026
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-