This vmsr submission is part of a retrospective data remediation review.The included events represent 7 malfunction events that medline became aware of during april 1, 2024 through june 30, 2024.Associated lot numbers 49123090002, 49123090005, and 74923090004.7 units of hcs4486 were returned for evaluation following reports of pin holes in the tubing.Medline quality received the samples and performed visual and functional evaluation.The samples arrived new in packaging and no visual defects were identified.The tubing was completely submerged in water while 5 liters of air was passed through the tubing.No leaks were identified during testing.The investigation summary documented the reported issue under item hcs4514 involving lot 49123090002 (5 units), lot 49123090005 (1 unit), and lot 74923090004 (1 unit).Based on the evaluation, the reported issue could not be confirmed and the cause of the reported condition could not be determined.7 devices were returned for evaluation.The devices were labeled for single use.No devices were identified as reprocessed or reused.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|