• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Zimmer Biomet, Inc. TPRLC 133 MP TYPE1 PPS SO 17.0; Prosthesis, Hip

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Zimmer Biomet, Inc. TPRLC 133 MP TYPE1 PPS SO 17.0; Prosthesis, Hip Back to Search Results
Catalog Number 51-106170
Medical Device Problem Codes Delivered as Unsterile Product (1421); Device Damaged Prior to Receipt by User (1570)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/13/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that damaged sterile packaging was identified during inspection of circulated stock items.No additional information available for the reported event.
 
Additional Manufacturer Narrative
(b)(4).G2: foreign ¿ japan.Visual examination of the returned product identifies that all the devices had creasing and tearing on the pouches at the pre-tear.Pouch seal width measured less than 6mm for all devices and are not acceptable.No other damage was noted during visual evaluation.Review of the device history records identified no deviations or anomalies during manufacturing.The root cause of the reported event can be attributed to transit damage and a packaging design issue.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TPRLC 133 MP TYPE1 PPS SO 17.0
Common Device Name
Prosthesis, Hip
Manufacturer (Section D)
Zimmer Biomet, Inc.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer (Section G)
Zimmer Biomet, Inc.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
+1(574)526-0384
MDR Report Key50370290
Report Number0001825034-2026-02454
Device Sequence Number5765607
Product Code KWA
UDI-Device Identifier00880304514416
UDI-Public(01)00880304514416(17)320711(10)7287915
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K110400
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 08/21/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Catalogue Number51-106170
Device Lot Number7287915
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer04/09/2026
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date08/21/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/11/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-