• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Zimmer Biomet, Inc. GLENOID IMPACTOR; Prosthesis, shoulder, semi-constrained, metal/polymer cemented

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Zimmer Biomet, Inc. GLENOID IMPACTOR; Prosthesis, shoulder, semi-constrained, metal/polymer cemented Back to Search Results
Catalog Number SBGL5001
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/24/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported during the procedure, it was discovered the instrument had a cracked tip that was about to break.The proper surgical technique was utilized.There was no patient injury or delay in procedure.Attempts have been made and no further information is available.
 
Additional Manufacturer Narrative
(b)(4).Multiple mdr reports were filed for this event.D10: concomitant medical products item: sbgl2000 lot: 64893257.The product has been received by zimmer biomet and the investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.If any further information is found that would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GLENOID IMPACTOR
Common Device Name
Prosthesis, shoulder, semi-constrained, metal/polymer cemented
Manufacturer (Section D)
Zimmer Biomet, Inc.
1800 w. center st.
warsaw IN 46580
Manufacturer (Section G)
Zimmer Biomet, Inc.
1800 w. center st.
warsaw IN 46580
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
+1(574)526-0384
MDR Report Key50371504
Report Number0001822565-2026-02913
Device Sequence Number18635689
Product Code KWS
UDI-Device Identifier00889024574854
UDI-Public(01)00889024574854(10)66153496
Combination Product (Y/N)N
Initial Reporter StateMT
Initial Reporter CountryUS
PMA/510(K) Number
K191814
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 08/21/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberSBGL5001
Device Lot Number66153496
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer08/10/2026
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date08/21/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
See H11
Patient SexMale
-
-