Clinical information: crd_1059 - vista nova study, patient site id: site id- (b)(6).It was reported that on (b)(6) 2026, a 27mm navitor vision valve was chosen for implant using a large flexnav delivery system.There wasn't calcification present extending beneath the aortic annular plane in the interventricular septum.The activated clotting levels were 183,189s and heparin was given.A pre-implant balloon valvuloplasty (bav) was done with a 20mm non-abbott balloon.After pre-bav, the patient had a left bundle branch block (lbbb).After multiple re-sheath (5 full re-sheath) due to implant too low, the valve got removed and a new valve was used.A new 27mm navitor vision valve and large flexnav delivery system were chosen.The final implant depth was 5.03mm on non-coronary cusp (ncc) and 7.32mm at left coronary cusp (lcc).During hospitalization, patient experienced atrial fibrillation with bradycardia and lbbb.A decision was made to implant a temporary pacemaker on (b)(6) 2026.The patient was reported discharged.
|